Attenuation correction of PET images with interpolated average CT for thoracic tumors
Huang T.-C.1,2; Mok G.S.P.3; Wang S.-J.1,4; Wu T.-H.1; Zhang G.5
2011-04-21
Source PublicationPhysics in Medicine and Biology
ISSN00319155 13616560
Volume56Issue:8Pages:2559-2567
Abstract

To reduce positron emission tomography (PET) and computed tomography (CT) misalignments and standardized uptake value (SUV) errors, cine average CT (CACT) has been proposed to replace helical CT (HCT) for attenuation correction (AC). A new method using interpolated average CT (IACT) for AC is introduced to further reduce radiation dose with similar image quality. Six patients were recruited in this study. The end-inspiration and -expiration phases from cine CT were used as the two original phases. Deformable image registration was used to generate the interpolated phases. The IACT was calculated by averaging the original and interpolated phases. The PET images were then reconstructed with AC using CACT, HCT and IACT, respectively. Their misalignments were compared by visual assessment, mutual information, correlation coefficient and SUV. The doses from different CT maps were analyzed. The misalignments were reduced for CACT and IACT as compared to HCT. The maximum SUV difference between the use of IACT and CACT was ∼3%, and it was ∼20% between the use of HCT and CACT. The estimated dose for IACT was 0.38 mSv. The radiation dose using IACT could be reduced by 85% compared to the use of CACT. IACT is a good low-dose approximation of CACT for AC. © 2011 Institute of Physics and Engineering in Medicine.

DOI10.1088/0031-9155/56/8/014
URLView the original
Indexed BySCIE
Language英語English
WOS Research AreaEngineering ; Radiology, Nuclear Medicine & Medical Imaging
WOS SubjectEngineering, Biomedical ; Radiology, Nuclear Medicine & Medical Imaging
WOS IDWOS:000289480700015
Scopus ID2-s2.0-79953722817
Fulltext Access
Citation statistics
Document TypeJournal article
CollectionFaculty of Health Sciences
DEPARTMENT OF ELECTRICAL AND COMPUTER ENGINEERING
Corresponding AuthorWu T.-H.
Affiliation1.Department of Biomedical Imaging and Radiological Sciences, National Yang Ming University, Taiwan
2.Department of Biomedical Imaging and Radiological Science, China Medical University, Taiwann
3.Department of Electrical and Electronics Engineering, Faculty of Science and Technology, University of Macau, Macau
4.Department of Nuclear Medicine, Taipei Veterans General Hospital, Taiwan
5.Department of Radiation Oncology, Moffitt Cancer Center, Florida, USA
Recommended Citation
GB/T 7714
Huang T.-C.,Mok G.S.P.,Wang S.-J.,et al. Attenuation correction of PET images with interpolated average CT for thoracic tumors[J]. Physics in Medicine and Biology, 2011, 56(8), 2559-2567.
APA Huang T.-C.., Mok G.S.P.., Wang S.-J.., Wu T.-H.., & Zhang G. (2011). Attenuation correction of PET images with interpolated average CT for thoracic tumors. Physics in Medicine and Biology, 56(8), 2559-2567.
MLA Huang T.-C.,et al."Attenuation correction of PET images with interpolated average CT for thoracic tumors".Physics in Medicine and Biology 56.8(2011):2559-2567.
Files in This Item:
There are no files associated with this item.
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[Huang T.-C.]'s Articles
[Mok G.S.P.]'s Articles
[Wang S.-J.]'s Articles
Baidu academic
Similar articles in Baidu academic
[Huang T.-C.]'s Articles
[Mok G.S.P.]'s Articles
[Wang S.-J.]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[Huang T.-C.]'s Articles
[Mok G.S.P.]'s Articles
[Wang S.-J.]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.